Disinfection Temperature Bands

Disinfection PeriodDisinfection TemperatureMaximum Temperature Allowed
10 minutes at
65°C
70°C
3 minutes at
73°C
78°C
1 minute at
80°C
85°C
0.2* minutes at
90°C
95°C
*BS 2745 Part 1 indicates that exposure time at 90°C is 1 second however this is too short for reliable measurement so 12 seconds (0.2 minutes) should be used.

Generic Instrument Washer Disinfector Cycle

Generic Instrument Washer Disinfector Cycle

 

 

 

 

 

 

Thermometric Test: Instrument Washer Disinfector

Thermometric TestThermocouple probes should be placed in the following positions:

  • A load item at each level of the load carrier
  • The area known to be slowest to reach temperature
  • The area known to be the quickest to reach temperature
  • Adjacent to the automatic control centre
  • Adjacent to the recorder sensor, if a multi-compartment machine one probe adjacent to each recorder sensor

  • Temperatures are to be within +/- 2°C for the required time.

    On completion of the cycle a check must be made to confirm all test probes remain in the correct positions. Due to the variation of instrument washer disinfectors and loads, advice may be required to determine if the results obtained are satisfactory.