Disinfection Temperature Bands
| Disinfection Period | Disinfection Temperature | Maximum Temperature Allowed |
|---|---|---|
| *BS 2745 Part 1 indicates that exposure time at 90°C is 1 second however this is too short for reliable measurement so 12 seconds (0.2 minutes) should be used. | ||
Generic Instrument Washer Disinfector Cycle

Thermometric Test: Instrument Washer Disinfector
| Thermometric Test | Thermocouple probes should be placed in the following positions: Temperatures are to be within +/- 2°C for the required time. On completion of the cycle a check must be made to confirm all test probes remain in the correct positions. Due to the variation of instrument washer disinfectors and loads, advice may be required to determine if the results obtained are satisfactory. |